Zonmedical
Chemistry Analyser KCA-A201

Chemistry Analyser KCA-A201

Chemistry Analyser KCA-A201 is a high-performance fully automated benchtop clinical chemistry analyzer. Random access based automation system increases test productivity and accuracy. System incorporates cuvette washing station and contact-free reaction design for better hygiene practices.

specifications

Type

Fully Automated Random Access System

Throughput

Reagent 1 : 400 test/hr
Reagent 2 : 600 test/hr
With ISE : 400 test/hr

Analytical Methods

End-point, Multi-standard, Immuno-turbid metric, 1-2 reagent, Reagent/serum blank, etc.

Assay Modules

80 modules

Sample Slots

80

Sample Volume

1µL - 100µL, (0.1µL/step)

Reagent Slots

2 x 40

Reagent Volume

R1 : 1µL - 400µL, (1µL/step)
R2 : 1µL - 400µL, (1µL/step)
1µL - 500µL, (0.1µL/step) [Min. 50 µL/test on request]

Probe Handling

Liquid level sensor, Collision sensor, Teflon coating and Automatic washing station.

Washing Reaction Cuvettes

8 channel Auto-washing

Reaction Time

0 s – 999 s

Alarms

Auto-alarm for crucial events (collision, reagent/ sample insufficiency, full waste container, etc.)

Calibration

Linear, Non-linear, Multi-point, K factor

Lamp

Halogen Lamp
Wavelength : 340 nm – 810 nm

Absorbance Resolution

0.0001 Abs

Absorbance Accuracy

± 0.0003 Abs ; ( 0 Abs – 2.5 Abs)

Repeatability

≤ 2 %

Quality Control

Levy-Jennings quality control at 3 levels

Power Supply

220 V / 50 Hz or 110 V / 60 Hz

Dimension

1000 x 850 x 700 mm

Weight

137 Kgs

features

  • Random access automation system
  • Up to 80 assay modules
  • Intelligent multi-tasking operation
  • Standard/ Control/ STAT modes
  • Auto-alarms with Levy Jennings QC
  • Tele-diagnostic network interface
  • Bar code sample scanning (Optional)

applications

  • Used in small to medium sized pathological diagnostic laboratories with wide range of applications including determination of the concentration of enzymes, substrates, electrolytes, specific proteins, drugs of abuse, and therapeutic drugs in samples for diagnosis of toxicological effect on renal, cardiac, and liver function.

catalog