Patient range : Adult, pediatric, neonate
NIBP Measurement method : Oscillometric
NIBP Measurement range : 0 kPa – 38.67 kPa (0 mmHg – 290 mmHg)
NIBP Measurement mode : Arm Measurement
Temperature Range : 37°C to 42°C
Temperature Accuracy : ± 0.5°C
Warm-up Time : < 45 s (20°C to 36°C)
Power Consumption : 250 VA
Temperature Range : 37°C to 42°C
Temperature Accuracy : ± 1.0°C
Warm-up Time : < 45 s (20°C to 36°C)
Power Consumption : 80 VA
Display : 12.1” LCD display
Ultrasound frequency : 1024 (1 ± 10%) kHz
Max output intensity : < 10 MW/cm2
FHR measurement range : 30 bpm ~ 240 bpm
Display : 7-inch screen
Waveform : Biphasic
Manual with Sync mode : 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 joules
AED mode : Auto analyze and charge x 3 with programmable auto energy level selection.
Waveform : Biphasic
Energy Sequence : 150 J, 200 J
Analysis Time : 9 seconds
Audio Indications : 20 distinct voice instructions
Display : OLED 160 x 128 dot matrix display
Storage : Built-in TF card storage for up to 2 GB
Standard Sensitivity : 10 mm/mV ± 5%
Sensitivity Shift : ≤ ± 10%
Control System : Air-temp control, Skin-temp control & Humidity control
Display : LED display
Air Temperature Control range : 25°C ~ 37°C (Override : 37°C ~ 39°C)
Baby Temperature Control range : 25°C ~ 37°C (Override : 37°C ~ 39°C)
Control System : Air-temp control & Skin-temp control
Display : LED display
Air Temperature Control range : 25°C ~ 37°C (Override : 37°C ~ 39°C)
Baby Temperature Control range : 25°C ~ 37°C (Override : 37°C ~ 39°C)
Control System : Air-temp control
Display : LED display
Air Temperature Control range : 25°C ~ 37°C
Baby Temperature Control range :
Channel : Single channel infusion
Volume (VTBI) : 1 ml ~ 9999.9 ml ( 0.1 ml increment)
Accuracy : Flow rate: ≤ ± 3 % Mechanical: ≤ ± 2 %
KVO Rate : 1 ml/h (1 ml/h ~ 300 ml/h) 3 ml/h (greater than 300 ml/h)
Channel : Single channel infusion
Infusion Volume : 0 ml ~ 9999 ml ( 0.1ml increment)
Infusion Rate : 0.01 ml/h ~ 1200 ml/h (0.01ml/h increment)
KVO Rate : 0.01 ml/h – 5 ml/h
Tidal Volume : 20 ml ~ 500 ml
Standard Ventilation Mode : VCV, SIMV, PSV, PCV, CPAP, Manual, SIGH
Application : Adult/ Pediatric Invasive
Pressure Range : 5 mbar ~ 80 mbar
Tidal Volume : 10 ml ~ 300 ml
Standard Ventilation Mode : CMV, IMV, CPAP, Manual
Application : Pediatric
Inspiratory Flow : 3 L/min ~ 20 L/min
Tidal Volume : 20 ml ─ 2500 ml
Configurations : Trolley or Ceiling Pendant
Respiratory Rate : 1 bpm ─ 40 bpm (V-SIMV/P-SIMV/PRVC-SIMV) 4 bpm ─ 100 bpm ( VCV/PCV/PRVC)
Psupport : 0 kPa ─ 6 kPa
Control System : Microprocessor based temperature control system
Control Modes : Preheat, Manual and Skin-temp mode
Manual : Heater output : 0-100%, 10% increments
Skin-temp : Control range : 25 °C ~ 37 °C
Vital Parameters :
| Type A | Type B | Type C |
|---|---|---|
| 4 Parameters (NIBP,SpO2, PR, Temp) | 3 Parameters (SpO2, PR, Temp) | 3 Parameters (NIBP, SpO2, PR) |
Display : 240 x 320 dot matrix colour TFT
NIBP Measurement range : Adult: 30-255 mmHg Paediatric: 30-160 mmHg
NIBP Mode : Auto, Manual, STAT
Vital Parameters : 3 Parameters (SpO2 (with PR), NIBP, & 1-channel TEMP)
NIBP Measurement range : Systolic Adult 50 ~ 265 mmHg Child 25 ~ 135 mmHg Diastolic Adult 15 ~ 220 mmHg Child 12 ~ 110 mmHg MAP Adult 25 ~ 245 mmHg Child 18 ~ 120 mmHg
NIBP Mode : Manual, Auto
NIBP Parameters : Sys, Dia, MAP
Standard Parameter : 5‐lead ECG, RESP, HR, NIBP, SpO2(F‐5), 1‐TEMP, PR
Patient Groups : Neonate, Pediatric, Adults
Trace : 9 waveform (ECG,SpO2, RESP)
ECG Lead Mode : 5–lead (R, L, F, N, C) for leads I, II, III, V, aVR, aVF, aVL
Light Source : 8 Blue LED lights
Wavelength : 450 to 475nm
LED Lamp-life : 1,00,000 hours
Intensity Control : Low : 25 ~ 35 μW /cm2/nm High : 35 ~ 55 μW /cm2/nm
Type : Adult
Stroke volume : 800 ml
Resuscitator volume : 1630 ml
Bag reservoir volume : 2000 ml
Type : Pediatric
Stroke volume : 350 ml
Resuscitator volume : 680 ml
Bag reservoir volume : 1600 ml
Type : Neonatal
Stroke volume : 150 ml
Resuscitator volume : 250 ml
Bag reservoir volume : 600 ml
KVO Rate : 0.1 – 5ml/h (steps : 0.1ml/h)
Injection Compliance : Adaptable for 5 syringe sizes
Preset Volume Range : 0.1 ml – 1999.9 ml
Occlusion System : 3-level alarming system
Channel : Single channel infusion
Volume (VTBI) : 0 ml – 9999 ml
Injection Compliance : Adaptable for 6 syringe sizes
Accuracy : ≤ ±2 %
Channel : Dual channel infusion
Volume (VTBI) : 0.1 ml – 1999.9 ml
Injection Compliance : Adaptable for 10 syringe sizes
Accuracy : ≤ ± 3 %
Tidal Volume : 0 ml – 1500 ml
Gas Supply : O2, 0.28 to 0.6 MPa
Ventilation Modes : A/C, SIMV, SPONT, Manual, VCV+SIGH
Respiratory Rate : 4 bpm – 40 bpm (SIMV) 4 bpm – 99 bpm (A/C)